Description
Feature:
1.Built-in circulating 3D constant temperature technology
Temperature stability time is less than 30 mins, ensuring the accuracy and reliability of test data.
2.The world’s leading full-range series of ion chromatographic columns
High efficiency, large capacity of the columns for detecting ions of varied compositions.
3. Self-Regenerating Electrolytic Micro-membrane Suppressor
High pressure resistance, small dead volume, highly responsive to signals.
4. Auto-range Conductivity Detector
It can directly detect the signal from ppb to ppm without adjusting the range; Only one conductivity detector can detect anions and cations.
5. Observatory intelligent workstation
Integrated control; With intelligent start-up, shutdown and maintenance functions; Compatible with a variety of instruments; Customized images.
Technical Parameter:
Model | BK-IC100 | |
IC Pump
|
Type |
High-pressure and low-pulse two-piston tandem advection pump |
Maximum Pressure |
35 MPa (PEEK) |
|
Pressure Display
Accuracy |
≤0.1MPa |
|
Flow Stability |
(0.2-0.5) mL/min≤3%;
(0.5-1.0) mL/min≤2%; >1.0mL/min≤2%; |
|
Allowable Deviation of Flow | (0.2-0.5) mL/min , ±5%;
(0.5-1.0) mL/min, ±3%; >1.0mL/min, ±2%; |
|
Flow Range | 0.001~9.999mL/min | |
Pressure Pulse | ≤0.5% | |
Flow Precision | ≤0.1% | |
Flow Accuracy | ≤0.1% | |
Manual Sample Injector |
Maximum Pressure | 35MPa |
External Diameter of the Tube | 1/16″ | |
Contact Material of the Rotor | PEEK | |
Contact Material of Medium | PEEK/Ceramics | |
Column Heater
|
Operating Temperature Range | 20~60℃ (68~140℉) |
Controlling Temperature Accuracy | ±0.01℃ | |
Allowable Deviation of Column Heater’s Temperature | ±1℃ | |
Temperature Stability | ≤0.05℃/h | |
Conduction
Detection System
|
Type | Constant temperature
auto-range conductivity detector |
Cell Volume
|
≤0.8μL | |
Detection Range | 0~35000μS/cm | |
Detection Resolution | ≤0.0020nS/cm | |
Output Voltage | -6000~+6000mv(adjustable) | |
Electronic Noise | 0.02nS | |
Baseline Noise | ≤0.001μS/cm | |
Baseline Drift | ≤0.02μS | |
Operating Temperature Range | Room temperature +5℃~60℃ (41 ~140℉) | |
Controlling Temperature Accuracy | ±0.01℃ | |
Temperature Compensation | 1.7 %/℃ | |
Maximum Pressure | 10.0MPa | |
Linear Range | ≥103 | |
Instrument Linearity | ≥0.999 | |
Quantitative Repeatability | ≤1.0% | |
Qualitative Repeatability | ≤0.5% | |
Minimum Detectable Concentration | Cl-≤0.0005ug/mL;
Li+≤0.001ug/mL; BrO3-≤0.001ug/mL; |
|
Flow System
|
Plastic Flow Path | Made of PEEK materials |
Six-way Valve | PEEK material, pressure 5000psi; Independent automatic collecting and flow function. | |
Suppressor | Type | Self-Regenerating electrolytic micro-membrane suppressor |
Maximum Pressure | 6.0MPa | |
Dead Volume | <50μL | |
Power Supply | 220V 50/60Hz(Standard);110V 50/60Hz (Optional);150W | |
External Size(L*W*H*) | 325*400*500mm | |
Net Weight | 21kg |